The inner surface of a burette must be thoroughly cleaned in order
that the liquid as drawn out may drain away completely, without
leaving drops upon the sides. This is best accomplished by treating
the inside of the burette with a warm solution of chromic acid in
concentrated sulphuric acid, applied as follows: If the burette is of
the "plain" type, first remove the rubber tip and force the lower
end of the burette into a medium-sized cork stopper. Nearly fill the
burette with the chromic acid solution, close the upper end with a
cork stopper and tip the burette backward and forward in such a way
as to bring the solution into contact with the entire inner surface.
Remove the stopper and pour the solution into a stock bottle to be
kept for further use, and rinse out the burette with water several
times. Unless the water then runs freely from the burette without
leaving drops adhering to the sides, the process must be repeated
(Note 1).
If the burette has a glass stopcock, this should be removed after
the cleaning and wiped, and also the inside of the ground joint. The
surface of the stopcock should then be smeared with a thin coating of
vaseline and replaced. It should be attached to the burette by means
of a wire, or elastic band, to lessen the danger of breakage.
Fill the burettes with distilled water, and allow the water to run out
through the stopcock or rubber tip until convinced that no air
bubbles are inclosed (Note 2). Fill the burette to a point above the
zero-point and draw off the water until the meniscus is just below
that mark. It is then ready for calibration.
[Note 1: The inner surface of the burette must be absolutely clean if
the liquid is to run off freely. Chromic acid in sulphuric acid is
usually found to be the best cleansing agent, but the mixture must be
warm and concentrated. The solution can be prepared by pouring over a
few crystals of potassium bichromate a little water and then adding
concentrated sulphuric acid.]
[Note 2: It is always necessary to insure the absence of air bubbles
in the tips or stopcocks. The treatment described above will usually
accomplish this, but, in the case of plain burettes it is sometimes
better to allow a little of the liquid to flow out of the tip while it
is bent upwards. Any air which may be entrapped then rises with the
liquid and escapes.
If air bubbles escape during subsequent calibration or titration, an
error is introduced which vitiates the
|