FREE BOOKS

Author's List




PREV.   NEXT  
|<   124   125   126   127   128   129   130   131   132   133   134   135   136   137   138   139   140   141   142   143   144   145   146   147   148  
149   150   151   152   153   154   155   156   157   158   159   160   161   162   163   164   165   166   167   168   169   170   171   172   173   >>   >|  
(6) L As T and t_{2} are taken with the same thermometer under the same set of conditions, any error in the reading of the thermometers will be approximately the same for the temperatures T and t_{2} and the above method therefore corrects for both the radiation and thermometer errors. The theoretical readings for dry steam, where there are no losses due to radiation, are obtainable from formula (5) by letting x = 0 and solving for t_{2}. The difference between the theoretical reading and the normal reading for no moisture will be the thermometer and radiation correction to be applied in order that the correct reading of t_{2} may be obtained. For any calorimeter within the range of its ordinary use, such a thermometer and radiation correction taken from one normal reading is approximately correct for any conditions with the same or a duplicate thermometer. The percentage of moisture in the steam, corrected for thermometer error and radiation and the correction to be applied to the particular calorimeter used, would be determined as follows: Assume a gauge pressure in the trial to be 180 pounds and the thermometer reading to be 295 degrees. A normal reading, taken in the manner described, gives a value of T = 303 degrees; then, the percentage of moisture corrected for thermometer error and radiation is, 0.47(303 - 295) x = ---------------- 845.0 = 0.45 per cent. The theoretical reading for dry steam will be, 1197.7 - 1150.4 - 0.47(t_{2} - 212) 0 = ------------------------------------ 845.0 t_{2} = 313 degrees. The thermometer and radiation correction to be applied to the instrument used, therefore over the ordinary range of pressure is Correction = 313 - 303 = 10 degrees The chart may be used in the determination of the correct reading of moisture percentage and the permanent radiation correction for the instrument used without computation as follows: Assume the same trial pressure, feed temperature and normal reading as above. If the normal reading is found to be 303 degrees, the correction for thermometer and radiation will be the theoretical reading for dry steam as found from the chart, less this normal reading, or 10 degrees correction. The correct temperature for the trial in question is, therefore, 305 degrees. The moisture corresponding to this temperature and 180 pounds gauge pressure will be found from the char
PREV.   NEXT  
|<   124   125   126   127   128   129   130   131   132   133   134   135   136   137   138   139   140   141   142   143   144   145   146   147   148  
149   150   151   152   153   154   155   156   157   158   159   160   161   162   163   164   165   166   167   168   169   170   171   172   173   >>   >|  



Top keywords:

reading

 

thermometer

 

radiation

 

degrees

 
correction
 
normal
 

moisture

 

theoretical


pressure

 

correct

 

applied

 

percentage

 

temperature

 

corrected

 

approximately

 

conditions


instrument

 
ordinary
 

calorimeter

 

Assume

 
pounds
 
manner
 

Correction

 

computation


question

 

permanent

 

determination

 
obtained
 

readings

 

errors

 

obtainable

 
losses

corrects

 
method
 

temperatures

 

thermometers

 
formula
 
duplicate
 

solving

 

letting


difference

 

determined