FREE BOOKS

Author's List




PREV.   NEXT  
|<   79   80   81   82   83   84   85   86   87   88   89   90   91   92   93   94   95   96   97   98   99   100   101   102   103  
104   105   106   107   108   109   110   111   112   113   >>  
f the analyzing prism, H, which it fits as well, but which must never be touched, as the adjustment of the instrument would be seriously disturbed by moving it. With the key on the proper nipple it is turned one way or the other until the field is neutral. Unless the deviation of the zero be greater than 0.5 deg., it will not be necessary to use the key, but only to note the amount of the deviation, and for this purpose the observer must not be content with a single setting, but must perform the operation five or six times, and take the mean of these different readings. If one or more of the readings show a deviation of more than 0.3 deg. from the general average, they should be rejected as incorrect. Between each observation the eye should be allowed 10 to 20 seconds of rest. The "setting" of the zero having been performed as above, the determination of the accurate adjustment of the instrument by means of the "control" quartz plates is proceeded with. Three such plates will be furnished with each polariscope, which have "sugar values" respectively approximating 80 deg., 90 deg., and 100 deg.. These values may vary with the temperature, and tables are furnished with them which give their exact value at different temperatures, from 10 deg. to 35 deg. C. One of these plates is placed in the instrument, and the field observed; it will be seen that the uniform appearance of the field is changed. The milled head is turned to the right until the exact point of neutrality is re-established, just as described above in setting the zero. The scale is read, the observation repeated, the reading taken again, and so on until five or six readings have been made. The average is taken, readings being rejected which show a divergence of more than 0.3, and the result corrected for the deviation of the zero point, if any was found, the deviation being added if it was to the left, and subtracted if to the right. If the adjustment of the instrument be correct, the result should be the value of the control plate used, as ascertained from the table, for the temperature of 20 deg.. Each of the three plates is read in the instrument in this way. A variation of 0.3 from the established values may be allowed for errors of observation, temperature, etc., but in the hands of a careful observer a deviation greater than this with one of the three plates, after a careful setting of the zero, shows that the instrument is not accurately adjusted.
PREV.   NEXT  
|<   79   80   81   82   83   84   85   86   87   88   89   90   91   92   93   94   95   96   97   98   99   100   101   102   103  
104   105   106   107   108   109   110   111   112   113   >>  



Top keywords:

instrument

 

deviation

 

plates

 

setting

 
readings
 

adjustment

 

temperature

 

observation

 

values

 

average


rejected
 

furnished

 
control
 
allowed
 

established

 

observer

 
greater
 

result

 
careful
 
turned

milled

 

changed

 

uniform

 

appearance

 
variation
 
errors
 

temperatures

 

observed

 

accurately

 

reading


repeated

 
corrected
 

adjusted

 

divergence

 

neutrality

 
correct
 

subtracted

 

ascertained

 
seconds
 

Unless


neutral

 

perform

 

operation

 
single
 

content

 

amount

 

purpose

 

nipple

 

proper

 

analyzing