The voltage generated by a thermo-couple of an electric pyrometer is dependent on the difference in temperature between its hot junction, inside the furnace, and the cold junction, or opposite end of the thermo-couple to which the copper wires are connected. If the temperature or this cold junction rises and falls, the indications of the instrument will vary, although the hot junction in the furnace may be at a constant temperature. A cold-junction temperature of 75 deg.F., or 25 deg.C., is usually adopted in commercial pyrometers, and the pointer on the pyrometer should stand at this point on the scale when the hot junction is not heated. If the cold-junction temperature rises about 75 deg.F., where base metal thermo-couples are used, the pyrometer will read approximately 1 deg. low for every 1 deg. rise in temperature above 75 deg.F. For example, if the instrument is adjusted for a cold-junction temperature of 75 deg., and the actual cold-junction temperature is 90 deg.F., the pyrometer will read 15 deg. low. If, however, the cold-junction temperature falls below 75 deg.F., the pyrometer will read high instead of low, approximately 1 deg. for every 1 deg. drop in temperature below 75 deg.F. With platinum thermo-couples, the error is approximately 1/2 deg. for 1 deg. change in temperature. CORRECTION BY ZERO ADJUSTMENT.--Many pyrometers are supplied with a zero adjuster, by means of which the pointer can be set to any actual cold-junction temperature. If the cold junction of the thermo-couple is in a temperature of 100 deg.F., the pointer can be set to this point on the scale, and the readings of the instrument will be correct. COMPENSATING LEADS.--By the use of compensating leads, formed of the same material as the thermo-couple, the cold junction can be removed from the head of the thermo-couple to a point 10, 20 or 50 ft. distant from the furnace, where the temperature is reasonably constant. Where greater accuracy is desired, a common method is to drive a 2-in. pipe, with a pointed closed end, some 10 to 20 ft. into the ground, as shown in Fig. 128. The compensating leads are joined to the copper leads, and the junction forced down to the bottom of the pipe. The cold junction is now in the ground, beneath the building, at a depth at which the temperature is very constant, about 70 deg.F., throughout the year. This method will usually control the cold-junction temperature within 5 deg.F. Where the greate
PREV.   NEXT  
|<   172   173   174   175   176   177   178   179   180   181   182   183   184   185   186   187   188   189   190   191   192   193   194   195   196  
197   >>  



Top keywords:
temperature
 

junction

 
thermo
 

pyrometer

 
couple
 

constant

 

instrument

 
approximately
 

pointer

 

furnace


pyrometers
 

ground

 

method

 

couples

 

actual

 
compensating
 

copper

 
COMPENSATING
 
correct
 

readings


removed

 

formed

 

material

 

greater

 

distant

 

building

 

beneath

 

greate

 

control

 

bottom


pointed
 

closed

 

common

 
desired
 

joined

 

forced

 

accuracy

 

adopted

 
commercial
 
heated

indications

 

electric

 
dependent
 

generated

 

voltage

 

difference

 

connected

 

opposite

 

inside

 

CORRECTION