The voltage generated by a thermo-couple of an electric pyrometer is
dependent on the difference in temperature between its hot junction,
inside the furnace, and the cold junction, or opposite end of the
thermo-couple to which the copper wires are connected. If the
temperature or this cold junction rises and falls, the indications
of the instrument will vary, although the hot junction in the furnace
may be at a constant temperature.
A cold-junction temperature of 75 deg.F., or 25 deg.C., is usually adopted
in commercial pyrometers, and the pointer on the pyrometer should
stand at this point on the scale when the hot junction is not heated.
If the cold-junction temperature rises about 75 deg.F., where base metal
thermo-couples are used, the pyrometer will read approximately 1 deg.
low for every 1 deg. rise in temperature above 75 deg.F. For example, if the
instrument is adjusted for a cold-junction temperature of 75 deg., and
the actual cold-junction temperature is 90 deg.F., the pyrometer will
read 15 deg. low. If, however, the cold-junction temperature falls below
75 deg.F., the pyrometer will read high instead of low, approximately
1 deg. for every 1 deg. drop in temperature below 75 deg.F.
With platinum thermo-couples, the error is approximately 1/2 deg. for
1 deg. change in temperature.
CORRECTION BY ZERO ADJUSTMENT.--Many pyrometers are supplied with
a zero adjuster, by means of which the pointer can be set to any
actual cold-junction temperature. If the cold junction of the
thermo-couple is in a temperature of 100 deg.F., the pointer can be
set to this point on the scale, and the readings of the instrument
will be correct.
COMPENSATING LEADS.--By the use of compensating leads, formed of
the same material as the thermo-couple, the cold junction can be
removed from the head of the thermo-couple to a point 10, 20 or 50
ft. distant from the furnace, where the temperature is reasonably
constant. Where greater accuracy is desired, a common method is
to drive a 2-in. pipe, with a pointed closed end, some 10 to 20
ft. into the ground, as shown in Fig. 128. The compensating leads
are joined to the copper leads, and the junction forced down to
the bottom of the pipe. The cold junction is now in the ground,
beneath the building, at a depth at which the temperature is very
constant, about 70 deg.F., throughout the year. This method will usually
control the cold-junction temperature within 5 deg.F.
Where the greate
|