FREE BOOKS

Author's List




PREV.   NEXT  
|<   68   69   70   71   72   73   74   75   76   77   78   79   80   81   82   83   84   85   86   87   88   89   90   91   92  
93   94   95   96   97   98   99   100   101   102   103   104   105   106   107   108   109   110   111   112   113   114   115   116   117   >>   >|  
ult to the intelligence of the physician to say that he cannot master a simple problem of electric testing involving the locating of one or more of five possibilities. It is simply a matter of memorizing five tests. It is repeated for emphasis that a commercial current reduced by means of a rheostat should never be used as a source of current for endoscopy with any kind of instrument, because of the danger to the patient of a possible "grounding" of the circuit during the extensive moist contact of a metallic endoscopic tube in the mediastinum. The battery shown in Fig. 8 should be used. The most frequent cause of trouble is the mistake of over-illuminating the lamps. _The lamp should not be over-illuminated to the dazzling whiteness usually used in flash lights_. Excessive illumination alters the proper perception of the coloring of the mucosa, besides shortening the life of the lamps. The proper degree of brightness is obtained when, as the current is increased, the first change from yellow to white light is obtained. Never turn up the rheostat without watching the lamp. _Testing for Electric Defects_.--These tests should be made beforehand; not when about to commence introduction. If the first lamp lights up properly, use it with its light-carrier to test out the other cords. If the lamp lights up, but flickers, locate the trouble before attempting to do an endoscopy. If shaking the carrier cord-terminal produces flickering there may be a film of corrosion on the central contact of the light carrier that goes into the carrier cord-terminal. If the lamp fails to show a light, the trouble may be in one of five places which should be tested for in the following order and manner. 1. The lamp may not be firmly screwed into the light-carrier. Withdraw the light-carrier and try screwing it in, though not too strongly, lest the central wire terminal in the lamp be bent over. 2. The light-carrier may be defective. 3. The cord may be defective or its terminals not tight in the binding posts. If screwing down the thumb nuts does not produce a light, test the light-carrier with lamp on the other cords. Reserve cords in each pair of binding posts are for use instead of the defective cords. The two sets of cords from one pair of binding posts should not be used simultaneously. 4. The lamp may be defective. Try another lamp. 5. The battery may be defective. Take a cord and light-carrier with lamp that lights up,
PREV.   NEXT  
|<   68   69   70   71   72   73   74   75   76   77   78   79   80   81   82   83   84   85   86   87   88   89   90   91   92  
93   94   95   96   97   98   99   100   101   102   103   104   105   106   107   108   109   110   111   112   113   114   115   116   117   >>   >|  



Top keywords:

carrier

 

defective

 
lights
 
binding
 

current

 

terminal

 

trouble

 

battery

 

contact

 

endoscopy


proper
 

screwing

 

central

 

rheostat

 
obtained
 
corrosion
 

flickers

 

properly

 

introduction

 

commence


locate

 

shaking

 

produces

 

attempting

 

flickering

 

produce

 

Reserve

 

terminals

 

simultaneously

 

manner


tested

 
places
 

firmly

 

screwed

 

strongly

 

Withdraw

 

brightness

 

source

 

commercial

 

reduced


instrument

 

extensive

 

circuit

 

grounding

 

danger

 

patient

 

emphasis

 
repeated
 

master

 

simple